3.5 Connection to Metrohm IC systems
Metrohm 838 Advanced Sample Processor
,
Installation
41
3.5.3 MIC-2 modular anionic system with chemical suppression
This system requires the
model 2.838.0010
of the 838 Advanced
Sample Processor.
The 838 Advanced Sample Processor is connected to an anionic IC
system with chemical suppression with 819 IC detector, 820 IC Separa-
tion Centre (2.820.0230), 818 IC pump, 833 IC Liquid Handling Unit and
830 IC Interface in accordance with
Figure 31
using the 6.2141.190 re-
mote cable and the 6.2134.100 RS232 cable.
819
838
820
818
830
PC
6.2134.100
6.2134.130
6.2115.070
6.2134.090
6.2125.090
6.2134.100
6.2141.190
833 Pump
6.2128.180
6.2134.040
Figure 31 MIC-2 interconnection with anionic system with
chemical suppression
838-specific settings in the «IC Net 2.3» program
Program for the 838 Advanced Sample Processor:
001 Ctrl
INIT
819
002 Move
sample
003 Lift
work
004 Ctrl
ZERO
1
005 Ctrl FILL A 1
006 Pump 240 s
007 Ctrl INJECT A 1
008 Start
DA
−
Initialise remote lines on 819
−
Move needle to sample position
−
Move lift with needle to working height
−
Trigger autozero on 819/1 IC Detector
−
Switch over injection valve on 820 "Fill"
−
Fill sample loop for 240 s with sample
−
Switch over injection valve on 820 to "Inject"
−
Start data recording