6 Operation and maintenance
947 Professional UV/VIS Detector Vario (2.947.00x0)
■■■■■■■■
55
Rinse again with ultrapure water and dry with a lint-free cloth.
4 Reinserting the lens
Insert a new inner seal
into the measuring cell. Make sure that
the inner seal lies flat and centered in the recess of the measuring
cell.
with its flat side facing inwards. Make sure
that the lens lies flat and centered in the recess of the measuring cell.
Insert a new outer seal
into the retaining screw
.
Reinsert the retaining screw
and tighten it with a slotted
screwdriver, size 5.
5 Cleaning the second lens
Repeat steps 2 to 4 on the opposite side of the cell holder.
6 Inserting the flow-through cell
Insert the flow-through cell such that the opening in the top right
corner is aligned with the cylinder screw on the cell block.
Figure 40
Inserting the flow-through cell
7 Screwing in the knurled screws
Screw in the knurled screws again.