![Metrohm 2.947.0010 Manual Download Page 53](http://html1.mh-extra.com/html/metrohm/2-947-0010/2-947-0010_manual_1779685053.webp)
■■■■■■■■■■■■■■■■■■■■■■
6 Operation and maintenance
947 Professional UV/VIS Detector Vario (2.947.00x0)
■■■■■■■■
45
6.3
Door
CAUTION
Incorrect detergents
The door is made of PMMA (poly(methyl methacrylate)). Property dam-
age to the door by using incorrect detergents.
■
Never clean the door with abrasive agents or solvents.
CAUTION
Instrument damage due to incorrect lifting
Property damage to the instrument due to incorrect lifting or moving.
■
Never hold the instrument by the door when lifting or moving it.
6.4
Replacing the lamps
CAUTION
Damage to the lamp
Reduced light transmission due to residues on the lamp surface. Dam-
age to the lamp due to burnt-on residues on the lamp surface.
■
Do not touch the lamp surface with your fingers!
■
If the lamp surface is contaminated, clean the lamp with alcohol and
a lint-free cloth.
■
Make sure the lamp surface is dry before using the instrument.
WARNING
Hot surface
Hot UV lamps and VIS lamps can lead to a risk of burns and fire.
■
Switch off the instrument and allow the lamp to cool before you
start working.
■
Do not place the hot lamp on flammable surfaces.