3.1 Base tray and bottle holder
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
2.945.0010 945 Professional Detector Vario – Conductometry
Accessories
■
3 mm hex key (6.2621.100)
1
Mount the base tray in such a way that the openings in the base tray
match exactly the screw threads in the bottom of the instrument.
2
Slide the washers onto the cylinder screws. Insert the cylinder screws
with the washers and tighten them with the 3 mm hex key.
3
Set the instrument back up on the base tray.
Stack other instruments in the required order. Mount the bottle holder
(6.2061.100) onto the topmost instrument on the stack
3.1.2.2
Removing/mounting the bottle holder
Remove the bottle holder if you want to install another instrument onto
the IC instrument.
Removing the bottle holder
Prerequisites
■
The instrument is switched off.
■
The bottle holder is cleared.
■
Drainage tubing is disconnected from the drainage tubing connection
of the bottle holder.
■
The capillaries are removed from the guide ducts between the instru-
ment and the bottle holder.
Accessories
■
3 mm hex key (6.2621.100)