3.6 Metrohm CO
2
Suppressor (MCS)
■■■■■■■■■■■■■■■■■■■■■■
30
■■■■■■■■
942 Extension Module Vario SeS/PP (2.942.0500)
CAUTION
If the MCS is not used, the inlet and outlet must be sealed with the
threaded stoppers (6.2744.220).
3.6.3
Installing adsorption cartridges
The aspirated air has to have the lowest CO
2
content possible in order for
the CO
2
to be removed from the eluent efficiently. In order to achieve this,
the air is aspirated through a CO
2
adsorption cartridge CW (6.2837.100).
Accessories
For this step, you need the following accessories:
■
Adsorption cartridge holder (6.2057.080) with two clips.
■
CO
2
adsorption cartridge CW (6.2837.100)
All of the parts are located in one accessory kit: Vario/Flex SeS
(6.5000.020).
Preparing the CO
2
adsorption cartridge CW
Prepare the CO
2
adsorption cartridge CW for use as follows:
1
Remove the protective cap from the tip of the CO
2
adsorption car-
tridge CW.
2
Attach the dust filter to the tip of the CO
2
adsorption cartridge CW.
3
Remove the label from the lid of the CO
2
adsorption cartridge CW.