3 Installation
942 Extension Module Vario SeS/PP (2.942.0500)
■■■■■■■■
13
3.3.2.2
Removing/mounting the bottle holder
Remove the bottle holder if you want to install another instrument onto
the IC instrument.
Removing the bottle holder
Prerequisites
■
The instrument is switched off.
■
The bottle holder is cleared.
■
Drainage tubing is disconnected from the drainage tubing connection
of the bottle holder.
■
The capillaries are removed from the guide ducts between the instru-
ment and the bottle holder.
Accessories
■
3 mm hex key (6.2621.100)
1
2
3
Figure 6
Removing the bottle holder
1
Remove the 2 covering stoppers.
2
Loosen the 2 cylinder screws with the 3 mm hex key and remove
them.
3
Remove the bottle holder.
Stack other instruments in the required order. Mount the bottle holder
(6.2061.100) onto the topmost instrument on the stack.
Mounting the bottle holder
Prerequisite