5.2 Maintenance by the user
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
■
Let go of the white lamp holder.
5 Closing the lamp compartment
■
Place the sealing plate onto the opening. Make sure that no
cables are being pinched.
■
Insert the four screws and tighten them in crosswise sequence
using the hex key.
■
Push the hex key into the guide rail on the lid.
■
Close the lid.
6 Calibrating the instrument
Recalibrate the instrument every time a lamp is replaced.
■
Switch on the instrument.
■
Wait for at least 2 hours while the instrument warms up.
■
Recalibrating the instrument
(see 8.105.8032 Tutorial Vision Air
Local and 8.105.8036 Tutorial Vision Air Network and Server).
5.2.5
Replacing the fan filter
Clean the fan filter at least once a month. If the instrument is operated in
a dusty or otherwise dirty environment, check the fan filter weekly or
twice a week as necessary.
The fan is on the rear of the instrument. The filter cover is attached to the
fan filter by means of four plastic latches.