5.2 Maintenance by the user
■■■■■■■■■■■■■■■■■■■■■■
22
■■■■■■■■
5.2.2
Cleaning the sample holder
Clean the sample holder using a lint-free cloth and ethanol.
NOTICE
Do not use aggressive solvents such as acetone. Aggressive solvents can
damage the surfaces of the instrument.
In the event of heavy soiling, one part of the sample holder can be
cleaned in a dishwasher. The sample holder consists of two parts. The
front part contains no electrical components. The rear part contains elec-
trical components. To separate the two parts from each other, unscrew
the two screws on the front part of the sample holder. Clean the front
part of the sample holder in a dishwasher.
Figure 5
Disassembling the sample holder
5.2.3
Cleaning the sample compartment
Dirt, dust and sample deposits in the sample compartment can influence
the analytical results.
Clean the sample compartment using a lint-free cloth and ethanol.
NOTICE
Do not use aggressive solvents such as acetone. Aggressive solvents can
damage the surfaces of the instrument.