5 Operation
899 Coulometer
■■■■■■■■
37
5.5
Control
Menu
▶
Control
In the dialog
Control
, the settings for the execution of a single determi-
nation or of one sample series are defined.
Sample table
If this parameter is activated, the sample data for a sample series can be
entered in a table
.
Selection
on
|
off
Default value
off
Autostart
If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the number specified
has been reached (see
Number of autostarts
).
Selection
on
|
off
Default value
off
NOTE
If you are working with a sample changer with oven module (e.g.
885
Compact Oven SC
), then the
Autostart
parameter must be set to
off
,
because the sample changer starts the titration.
Number of autostarts
This parameter is visible only when
Autostart
=
on
.
Number of automatic starts.
Input range
1 - 50
Selection
table
Default value
table
table
The number of automatic starts corresponds to the number of samples
in the sample table.