3.2 Carrying out titrations
■■■■■■■■■■■■■■■■■■■■■■
30
■■■■■■■■
870 KF Titrino plus
The result and the sample size of each determination are shown.
On/
Off
removes the selected determination from the statistics. The line
will then be marked with an asterisk (*), the statistics will be recalcu-
lated automatically.
7 Cancel the conditioning
■
Press
[STOP]
.
Conditioning is canceled and the dosing cylinder is filled.
3.2.4
Creating a titration method
Below you will find a description of how to create your own method for
water content determination. The 870 KF Titrino plus contains method
templates which are already configured except for a few parameters.
The water content of a sample should be determined by means of a triple
determination. If you have a printer connected, then a report with the
result and the curve should be printed out automatically at the end of the
determination.
Proceed as follows:
1 Load the method template
■
In the main dialog, select
Method
and confirm with
[OK]
.
The method table with the stored templates opens:
■
Select the template
KFT Ipol
.
■
In the function bar, select the
Load
function and confirm with
[OK]
.
The method template
KFT Ipol
is now loaded and is displayed in the
main dialog under
Method
.
Summary of Contents for 2.870.0010
Page 1: ...870 KF Titrino plus Installation and tutorial 8 870 8004EN...
Page 2: ......
Page 6: ......