3.18 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
1
3
2
3
4
5
6
7
8
2
1
Figure 30
MSM-HC – Connectors
1
Union nut
2
MSM-HC connecting piece 6.2835.010
3
Eluent inlet capillary
Labeled with
Eluent
.
4
Eluent outlet capillary
Labeled with
Detector
.
5
Rinsing solution inlet capillary
Labeled with
H2O
.
6
Rinsing solution outlet capillary
Labeled with
Waste
.
7
Regeneration solution outlet capillary
Labeled with
Waste
.
8
Regeneration solution inlet capillary
Labeled with
H2SO4
.
The PTFE capillaries firmly mounted on the MSM-HC are connected to the
other components of the IC system as follows:
CAUTION
As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.
Flattened capillaries can be shortened with the help of a capillary cutter.