6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
6 Troubleshooting
6.1
Problems and their solutions
Problem
Cause
Remedy
Marked drop in
pressure
Leak in the system.
Marked rise in pres-
sure
6.2821.120 inline filter
blocked.
MSM-HC – blocked.
■
.
Note: 6.2821.180 pump tubing connection
with filter must be used .
Conductivity detector
blocked.
■
Shorten capillary ends by several mm
■
Rinse detector against the normal flow
direction
(see Chapter 5.14.1, Page 105)
Guard column – blocked.
Separation column –
blocked.
■
Regenerate separation column
.
■
Replace separation column
ing and rinsing the separation column",
page 73)
Note: Samples should always be microfiltered
.
Injection valve – valve
blocked.
Have the valve cleaned (by Metrohm service
technicians).
Drift of the baseline
Thermal equilibrium not yet
attained.
ment with the column thermostat switched
on .
Leak in the system.