Metris Instruments TN418L1 Operating Instructions Download Page 1

Ref.No.

042007 

TN418L1 / TN418L1E

(with white LED Flashlight) 

Thermometer

 

Operating Instructions

 

 

 

 

The thermometer is a non-contact infrared thermometer, and always with Real-Time Maximum (MAX) data.   
Please remember to keep away from children and don’t use it for safety related applications. 
 

 

 

 

 

(Default Screen) 

 

ADD VALUE: 

o

C or 

o

Press Mode key (

2

) for 

o

C or 

o

F transferred. 

Backlight

 

LCD Backlight: always on. 

Laser

 

ClassII Laser: always enable while measuring. 

 

EMC/RFI

 

Readings  may  be  affected  if  the  unit  is  operated  within  radio  frequency  electromagnetic  field  strength  of 
approximately 3 volts per meter, but the performance of the instrument will not be permanently affected. 
* Note: Under the electromagnetic field of 3V/m from 350 to 550MHZ, the maximum error is 8°C (14.4°F). 

 

 

 

 

CAUTION 

1.

 

WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT. 

2.

 

USE EXTREME CAUTION WHEN OPERATING THE LASER. 

3.

 

NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.   

4.

 

KEEP OUT OF REACH OF ALL CHILDREN. 

STORAGE & CLEANING 

The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when 
cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the 

thermometer. Do not submerge any part of the thermometer. The thermometer should be stored at room temperature between –20 to 
+65°C (-4~149°F).

 

LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic messages as follows: 

 

 

‘Er2’  is  displayed  when  the  thermometer  is  exposed  to  rapid  changes  in  the  ambient  temperature.  ‘Er3’  is 
displayed when the ambient temperature exceeds 0

o

C(32°F) or +50

o

C (122°F).   

The  thermometer  should  be allowed  plenty  of  time  (minimum  30  minutes)  to  stabilize  to  the  working/room 
temperature. 

 

Error 5~9, for all other error messages it is necessary to reset the thermometer. To reset it, turn the instrument 

off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error 
message remains please contact the Service Department for further assistance. 

 

 

‘Hi’  or  ’Lo’  is  displayed  when  the  temperature  being  measured  is  outside  of  the 

measurement range.

 

BATTERIES

 

The thermometer incorporates visual low battery indication as follows: 

 

 

 

‘Battery OK’: measurements 

  are possible 

‘Battery Low’: battery needs to be   

replaced, measurements are still possible 

‘Battery Exhausted’: measurements are 

not possible 

  When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries. 

Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction. 
Dispose of used battery promptly and keep away from children. 

SPECIFICATION 

Measurement Range 

-60 to +500

 o

C (-76 to +932

 o

F) 

Operating Range 

0~50°C (32~122°F) 

Accuracy, (Tobj=15-35°C,Tamb=25°C) 

+/-1.0°C (1.8°F) 

Accuracy (Tamb=23 +/- 3°C) 

0~500°C:+/-2% of reading or 2°C (4°F) whichever is greater 
-60~0

 o

C: +/-(2°C+0.05/degree) 

Emissivity 

0.95 fixed 

Resolution (-9.9~199.9°C) 

0.1°C/0.1°F 

Response Time (90%) 

1sec 

Distance:Spot 

12:1 

Battery Life 

Typ.18, min 14 hours continuous use 

Dimensions 

46.0 x 143.0 x 184.8mm(1.81×5.63×7.28 inch) 

Weight 

240 grams(8.5 oz) including batteries (AAA*2pcs) 

** The thermometer will automatically shut off if left idle for more than 15sec. 

1. Simply aim the thermometer at the measure target with Lens (

4

) and press 

Meas. key (

3

) to display the surface temperature.     

 

2. During the measurement, the MAX reading will be displayed beside the 

 

icon. 

 

3. The Distance:Spot is 12:1. Please make sure the target area is within the field   

of view.

 

Reviews: