individual Single test results are stored in the Auto Sequence
®
result as if they were
programmed as independent measuring steps.
E.4
Creating / modifying an Auto Sequence
®
If creating a new Auto Sequence
®
from scratch, the first step (Header) and the last step (Result)
are offered by default. Measurement steps are inserted by the user.
Options:
Adding a measurement step
By double clicking on a Single test a new
measurement step will appear as the last of
measurement steps. It can also be dragged and
dropped on the appropriate position in the Auto
Sequence
®
.
Adding flow commands
Selected flow command can be dragged from the list
of Flow commands and dropped on the appropriate
place in any Auto Sequence
®
step.
Changing position of flow command
within measurement step
By a click on an element and use of
and
keys.
Viewing / changing parameters of
flow commands or single tests.
By a double click on the element.
Setting number of measurement
step repetitions
By setting a number in the
field
Right click
on the selected measurement step / flow command:
Copy
– Paste before
A measurement step / flow command can be copied and
pasted above selected location on the same or on another
Auto Sequence
®
.
Copy
– Paste after
A measurement step / flow command can be copied and
pasted under selected location on the same or on another
Auto Sequence
®
.
Delete
Deletes the selected measurement step / flow command.
E.5
Description of flow commands
Double click on inserted Flow Command opens menu window, where text or picture can be
entered, external signalling and external commands can be activated and parameters can be
set.
Flow commands Operation after end of test and Results screen are entered by default, others
are user selectable from Flow Commands menu.
Pause