MI 3250 MicroOhm 10A
Technical specification
36
8 Technical specifications
8.1 Resistance measurement
Test current
Resistance Range
Resolution
Accuracy
10
A
0000,0
… 2000,0
µΩ
0,1
µΩ
00,000
… 20,000
mΩ
1
µΩ
±(0,25% Rdg + 0,01%FS)
000,00
… 200,00
mΩ
10
µΩ
1 A
00,000
… 20,000
mΩ
1
µΩ
000,00
… 200,00
mΩ
10
µΩ
±(0,25% Rdg + 0,01%FS)
0,0000
… 2,0000
Ω
100
µΩ
100 mA
000,00
… 200,00
mΩ
10
µΩ
0,0000
… 2,0000
Ω
100
µΩ
±(0,25% Rdg + 0,01%FS)
00,000
… 20,000
Ω
1
mΩ
10 mA
0,0000
… 2,0000
Ω
100
µΩ
00,000
… 20,000
Ω
1
mΩ
±(0,25% Rdg + 0,01%FS)
000,00
… 200,00
Ω
10
mΩ
1 mA
00,00
… 20,00
Ω
10
mΩ
±(1% Rdg + 0,1%FS)
000,0
… 200,0
Ω
100
mΩ
±(1% Rdg + 0,25%FS)
0,000
… 2,000
kΩ
1
Ω
±(1% Rdg + 0,25%FS)
Table 8.1: Resistance measurement ranges and accuracy (reference conditions)
Test current accuracy ...............
±10 % (smoothed DC)
Test duration ............................. 1 s (single mode)
Test method .............................. 4-wire measurement
Note:
All data regarding accuracy is given for nominal (reference) environment
condition and forward and reverse measurements.
Inductive mode will introduce an undefined error if an EMF is present on the test
object.
The error in operating conditions could be at most the error for reference
conditions (specified in the manual for each function) +0,1% of measured value +
1 digit, unless otherwise specified in the manual for particular function.
Noise rejection (50/60 Hz) on potential leads P1 - P2:
Max. Noise
Level
Test Current
Additional
Error
10
A
1
A
100
mA
10
mA
1
mA
Resistance
Range
20
mV
RMS
50
mV
RMS
250
mV
RMS
2
mΩ
20
mΩ
200
mΩ
20
mΩ
200
mΩ
2
Ω
200
mΩ
2
Ω
20
Ω
2
Ω
20
Ω
200
Ω
20
Ω
200
Ω
2
kΩ
±10
dig
8.2 Measurement parameters
Test Current:
10
A
1
A
100 m
A
10 m
A
1 m
A
Max. Power Output:
20
W
2
W
0,2
W
20
mW
2
mW
Output Voltage:
3
V
DC
max.
1
V
DC
max.
Limits:
1
μΩ
… 2 k
Ω
Table 8.2: Measurement parameters