MI 3100 s, MI 3100 SE
Eurotest
EASI
Measurements
26
Keys:
UP / DOWN
Sets Isc value.
TEST
Confirms Isc value.
ESC
Exits back to settings menu.
Function selector
switch
Exits back to selected test / measurement function
without changes.
Short circuit current Isc in the supply system is important for selection or verification of
protective circuit breakers (fuses, over-current breaking devices, RCDs).
The default value of Isc factor (ksc) is 1.00. The value should be set according to local
regulative.
Range for adjustment of the Isc factor is 0.20
3.00.
4.3.6 Commander support (MI 3100 SE only)
The support for remote commanders can be
enabled or disabled in this menu.
Figure 4.9: Selection of commander support
Keys:
UP / DOWN
Selects commander option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function selector
switch
Exits back to selected test / measurement function
without changes.
Note:
Commander disabled
option is intended to disable the commander’s remote keys. In the
case of high EM interfering noise the operation of the commander can be irregular.
4.3.7 Communication (MI 3100 SE only)
In this menu the Bluetooth dongle A1436 can be
initialized and device for scanning barcodes can
be selected.
Figure 4.10:
Communication menu
Keys:
UP / DOWN
Selects option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function selector
switch
Exits back to selected test / measurement function
without changes.
Initialization of the Bluetooth dongle