MI 3155
Eurotest
XD
Auto Sequences
186
During the execution phase of an Auto
Sequence®, pre-programmed single tests are
carried out. The sequence of single tests is controlled by pre-programmed flow
commands.
After the test sequence is finished the Auto
Sequence® result menu is shown. Details of
individual tests can be viewed and the results can be saved to Memory organizer.
8.2.1
Auto Sequence® view menu
In the Auto
Sequence® view menu, the header and the single tests of selected Auto
Sequence® are displayed. The header contains Name, Short code and description of the Auto
Sequence®. Before starting the Auto Sequence®, test parameters / limits of individual
measurements can be changed.
Note:
Once fuse and RCD parameters
are changed in active Auto Sequence®, the new
settings are distributed through all single tests within the active Auto Sequence® and
stored for next use of the same Auto Sequence®.
8.2.1.1
Auto Sequence® view menu (Header is selected)
Figure 8.3: Screen organization in Auto
Sequence® view menu – Header selected
Legend
1
Short code
2
Auto Sequence
® name
3
Header
4
Single tests
5
Description
6
Control panel (available options)
Option
Starts the Auto
Sequence®.