MI 3152(H)
Eurotest
XC (2,5 kV)
Appendix B
– Profile notes
196
Modifications in chapters:
– Fault loop impedance and prospective fault current in system with RCD
– Auto test for TT earthing system
7.29 Z auto - Auto test for fast line and loop testing
15 mA added in the
I
N
parameter in
Test Parameters / Limits
section as follows:
I
N
Rated RCD residual current sensitivity
[10 mA, 15 mA, 30 mA, 100 mA,
300 mA, 500 mA, 1000 mA]
Nominal residual current (A,AC) ............. 10 mA, 15 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
RCD test current in relation to RCD type, nominal RCD current and multiplication factor
I
N
× 1/2
(mA)
I
N
× 1
(mA)
I
N
× 2
(mA)
I
N
× 5
(mA)
RCD I
I
N
(mA) AC A, F B, B+ AC
A, F B, B+ AC
A, F B, B+ AC A, F B, B+ AC A, F B, B+
10
5
3.5
5
10
20
20
20
40
40
50
100 100
15
7.5 5.3
7.5
15
30
30
30
60
60
75
150 150
30
15 10.5
15
30
42
60
60
84
120
150 212 300
100
50
35
50
100
141
200
200
282
400
500 707 1000
300
150 105
150
300
424
600
600
848
×
1500
×
×
500
250 175
250
500
707 1000 1000 1410
×
2500
×
×
1000
500 350
500 1000 1410
×
2000
×
×
×
×
×
×
×
. ............................................................ not applicable
. ............................................................ applicable
AC type................................................... sine wave test current
A, F types…… ........................................ pulsed current
B, B+ types ............................................. smooth DC current
Other technical specifications remain unchanged.
A.6 Profile Denmark (profile code ALAM)
DC offset for pulsed test current ............. disabled