MI 2142 AlphaPAT Measurements
25
CLASS
APPLIANCE
UNDER
TEST
II
Measurement of substitute leakage current of class II appliance
6.3.3 How to perform single substitute leakage current measurement
Select
Sub leakage
in
Single test menu
by using and cursor keys and press
ENTER key to confirm.
Sub leakage menu
is displayed.
Sub leakage menu
Measurement parameters are detailed in the lower left corner of the display. The
following parameters can be adjusted in this measurement:
high limit leakage value and
measurement duration.
Use and cursor keys to select the parameter you want to edit. Use and cursor
keys to adjust the selected parameter.
Press START key to perform measurement.
Examples of substitute leakage current measurement results
The measurement can be aborted by pressing START key at any time.
Actual result is shown on the display during measurement. After measurement time has
elapsed the last measured value is displayed, and PASS or FAIL indication appears on
the display.
Press START key to repeat the test or press ESC key to return to the
Single test
menu
.