MI 2142 AlphaPAT Technical specifications
14
3.4 Leakage current and apparent power
Differential leakage current readout
Range Resolution
Accuracy
0.00 mA
÷
19.99 mA
0.01 mA
±
(5 % of r 5 digits)
Pass levels ............... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.25 mA,
2.50 mA, 3.00 mA, 3.50 mA, 9.90 mA
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s
Output....................... test socket
Apparent power readout
Range Resolution Accuracy
0.00 kVA
÷
3.7 kVA
0.01 kVA
±
(10 % of r 3 digits)
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s
Output....................... test socket
3.5 Touch
leakage
current
Touch leakage current readout
Range Resolution Accuracy
0.00 mA
÷
1.99 mA
0.01 mA
±
(5 % of r 5 digits)
Pass levels ............... 0.25 mA, 0.50 mA, 1.00 mA
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s
Output....................... test socket,
test probe connector
Probe ........................ human body model with resistance of 2 k
Ω
3.6 Polarity
test
Test voltage .............. <50 V AC
Detects ..................... Pass, L-open, N-open, PE-open, L-N crossed, L-PE crossed, N-
PE crossed, L-N shorted, L-PE shorted, N-PE shorted, multiple
faults
Output....................... test socket, IEC cord