23
VARIOS
Please follow the instructions below to perform a calibration test:
1. Remove the VARIOS from the HYPROP Balance for calibration.
2. Replace the original measurement pan that came with the Balance.
3. Be sure nothing is placed on the Balance.
4. Press the
ZERO/TARE button on the Balance weight.
5. Press and hold the
FUNCTION button until S.A. CAL appears on the display.
6. Press the
FUNCTION and ZERO/TARE buttons at the same time.
7. Release the
FUNCTION and ZERO/TARE buttons at the same time.
8. Wait until
CAL. ON
appears on the display.
9. Turn the blue calibration knob on the right side of the Balance to
CAL..
VARIOS
Weight
calibration knob
HYPROP
Balance
Figure 24 Weight calibration knob (blue)
The internal weight will be activated. The adjustment runs automatically.
10. Turn the calibration knob back to
WEIGH when the display reads CAL. OFF.
Calibration is finished when the display reads
END.
The calibration can be verified by placing a 1 kg weight on the Balance. The Balance
should read within ±0.05 g of the weight.
2.7.3.3 DEFAULT SETTINGS FOR VARIOS MEASUREMENT
Find the HYPROP Balance settings by pressing the
FUNCTION button until
func
shows on the
display. METER recommends using the settings listed in
for measurements.