Specifications | 2-1
SPECIFICATIONS
The standard customer range is defined as full-scale ranges that fall
within the minimum and maximum allowable for a given Measuring
Cell. Calculations of minimum and maximum are based on atmospheric
pressure.
Methodology
Multiple Wavelength, high resolution, non-dispersive ultraviolet.
(Optional O
2
Sensor: Zirconium Oxide cell –
for use in General Purpose
applications/analyzers only
.)
Standard Ranges
Maximum ranges are based on Measuring Cell length (expressed in cm).
Minimum ranges are 1/50 of maximum ranges.
SO
2
SO
2
Range = 20 % SO
2
/ (Cell length), to a minimum of 250 PPM
NO
NO Range = 20 % NO / (Cell length), to a minimum of 300 PPM
NO
2
NO
2
Range = 40 % NO
2
/ (Cell length), to a minimum of 300 PPM
NO
x
NO
x
Range = 20 % NO
x
/ (Cell length), to a minimum of 300 PPM
H
2
S H
2
S Range @ 228 nm = 40 % H
2
S / (Cell length), to a minimum of
500 PPM
H
2
S Range @ 214 nm = 10 % H
2
S / (Cell length), to a minimum of
125 PPM
NH
3
NH
3
Range = 10 % NH
3
/ (Cell length), to a minimum of 125 PPM
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...