4-62 | Model 920 Multi-Gas Analyzer
Auto-Calibration
The Auto-Calibration function automatically performs an Auto-Zero, fol-
lowed by Auto-Span 1 and Auto-Span 2. To eliminate a function from the
sequence, set its timer to ‘
0
’. The sequence of events for the component is
the same as described for Auto-Zero/Auto-Span.
Auto-Span 2
and its associated timer (
Timer 2
) is used only when
the optional O
2
Sensor is used (
General Purpose applications/ana-
lyzers only
).
This function can be started manually, automatically on a timed basis, or
by a remote dry (potential-free) contact closure. Upon completion of the
calibration sequence, the Cal Status relay will continue to indicate that a
calibration is in process until the
SDelay
timer has expired.
Timed Start of Auto-Calibration
The analyzer automatically initiates the Auto-Calibration sequence on
a timed basis. The
ACal Timer
is the time interval between calibration
sequences. To set this function, press
F3 9
(CAL mode) and then enter a
value (hours) for the automatic-timed start of the Auto-Calibration.
If the
ACal Timer
is set to an interval other than ‘
0
’, the first Auto-
Calibration occurs after the
ACal
interval from the time the analyzer
is powered up or reset as long as no Fault conditions are present.
Subsequent Auto-Calibrations will be initiated at every
ACal
interval.
Example:
If the
ACal Timer
is set to ‘
3
’ and the analyzer is powered up or reset, the
first Auto-Calibration will occur at 3 hours if all temperatures are reached.
NOTE
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...