4-34 | Model 920 Multi-Gas Analyzer
CAL Mode – F2 Commands
Keystroke Term
Definition
F2 •
—
—
F2 –
—
—
F2 0
Man/Zero?
Allows you to manually Zero the analyzer. The analyzer Zero is automatically adjusted
based on the average reading during the
IntTime
duration after a
Yes
response only
if
AdjDisable
is set to ‘
0
’. The Zero gas must be introduced manually into the analyzer
sample system. Refer to “Manually Zeroing the Analyzer” in Chapter 3.
F2 y
Man/Spany? Allows you to manually Span the analyzer for each component, where ‘y’ = Manual Span
F2 1..F2 6
. The analyzer calibration is automatically adjusted based on the average read-
ings during the
IntTime
duration after a
Yes
response only if
AdjDisable
is set to ‘
0
’. The
Calibration (Span) gas must be introduced manually into the analyzer sample system.
Refer to “Manual Span” in this chapter for details about how to manually Span the analyzer.
F2 1 = Man/Span1?
F2 4 = Man/Span4?
F2 2 = Man/Span2?
F2 5 = Man/Span5?
F2 3 = Man/Span3?
F2 6 = Man/Span6?
F2 7
—
—
F2 8
Man/SpanO2? Allows you to manually Span the O
2
Sensor (if used –
General Purpose applications/
analyzers only
) at the analyzer. The sensor Calibration is automatically adjusted based on
the average reading during the
IntTime
duration after a
Yes
response only if
AdjDisable
is
set to ‘
0
’. Man/Span must be used to calibrate the sensor if Nitrogen is used for the Span
gas. The Calibration (Span) gas must be introduced manually into the analyzer sample sys-
tem. For systems with the O
2
Sensor, refer to the
O
2
Concentration Measurement Manual
Supplement
for details about how to manually Span the O
2
Sensor.
F2 9
Man/ZeroO2? Allows you to manually Zero (air point calibration) the O
2
Sensor (if used –
General
Purpose applications/analyzers only
) at the analyzer. The sensor Zero is automatically
adjusted based on the average reading during the
IntTime
duration after a
Yes
response
only if
AdjDisable
is set to ‘
0
’. Man/Zero must be used to calibrate the sensor if Nitrogen is
used for the Zero gas. The Zero gas must be introduced manually into the analyzer sample
system. For systems with the O
2
Sensor, refer to the
O
2
Concentration Measurement
Manual Supplement
for details about how to manually Zero the O
2
Sensor.
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...