3-42 | Model 920 Multi-Gas Analyzer
Manually Zeroing the Analyzer
A Manual Zero forces the analyzer to purge the sample system with Zero
gas which removes any residual contaminants in the sample system.
This mode is also used to set the timers to control the Auto-Zero. While
a Manual Zero is in progress, the RUN mode screen will display “
Z
” on
the top-left line to indicate that the analyzer has been
manually
forced to
Continuous Zero Gas Flow
mode.
All temperature zones must be at their operating Set Points before
performing a Manual Zero. If necessary, view
MAI
screen and check
the temperature zones to ensure that all zones are at their operating
temperatures.
This procedure assumes Instrument Air is used as the Zero gas.
Pressing
Esc
at any time during this procedure will abort the proce-
dure and return to CAL mode to normal display.
To manually Zero the analyzer:
1. Ensure that the Zero gas cylinder is turned on.
2. Before changing the Flow Control mode, record the current numeric
value (should be “
0
” –
Analyzer Control
mode). Change the
Flow
Control
mode to
Continuous Zero Gas Flow
mode, either “
10
” (
probe
tip
) or “1
10
” (
manual
).
3. From CAL mode initiate the Manual Zero. The User Interface will
prompt “
Man/Zero?
”. Observe the readings on the bottom line. When
the readings have stabilized at or near zero, press:
Ent
for Yes
The
IntTime
duration will count down from 15 seconds
(default if
IntTime
is ‘
0
’) to zero during which time
the readings are averaged if
AdjDisable
is set to ‘
0
’.
The Zero values will be adjusted automatically and the
screen will revert to CAL mode normal display.
Esc
for No
The function is aborted and the Zero values are not
adjusted.
Run the Manual Zero for 10 minutes at a flow rate of 2.5 L/min (0.08
SCFM). See “Setting the Zero Gas Flow Rate” in this chapter.
If the User Interface displays a “
w ZDrift/SpnErr
” alarm message,
restart the Manual Zero (enter
F2 0 Ent
).
NOTE
(FlowCtrl) CAL
F5 2
Del 10 Ent Ent
(Man/Zero?) CAL
F2 0
(MAI) RUN
F6 8
1..8
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...