8.1.
Thermal Insulation
The integration the FTC200-OEM into a custom housing is mandatory,
since the device should be prevented from draught and touch for specified
performance. It may not be exposed to airflow, e.g. from fans. The use of
additional insolation is optional, but maybe necessary for ambient
temperatures below 0°C. Do not use flammable material for insulation. In
case it is in contact to the metal body, the material used must be
temperature resistant up to 120°C. It may not cover the PCBs. We
recommend an insulation material that is heat resistant up to 170°C.
The device is equipped with a temperature fuse switching off at 110°C on
the surface of the stainless-steel body.
This device is specified for an ambient temperature at direct place of
operation between -10°C to 50°C. Please consider the waste heat of other
units nearby. They may not cause a temperature enhancement.
Do not mount the device directly on a surface to prevent strong thermal
contact. Use spacers with a length of 4mm that are temperature-resistant
up to 100°C.
9.
Communication with the FTC200-OEM
9.1.
Remote Control via Serial Communication