10.
Specifications
Dimensions with connectors;
weight
B=64mm, H=40mm,
D=75mm; ~ 350g
Power supply
24V DC (20V to 30V), 700mA
RS232 - Baudrate / Data
19200 / 8bit
Ambient temperature range
-10°C to 50°C, depending on
mounting and thermal insulation
Linearity
< 1% of range
Warm up time
Approx. 20min; 1h for small
ranges
Flow rate
10l/h-100l/h, recommended
60l/h to 80l/h
T90-time
<1sec at flow rate higher 60l/h
Noise
< 1% of smallest range
Drift at zero point
< 2% of smallest range per week
Repeatability
< 1% of range
Error due to change of ambient
temperature
< 1% of smallest range per 10°C
Error due to change of flow at
70l/h
< 1% of smallest range per
10l/h
Gas pressure (absolute)
2000kPa (20bar)
Error due to change of pressure
(above 800hPa)
< 1% of smallest range per
10hPa
Note: The values given above refer to H2 in N2, they may vary for other
gas pairs.