FTC200-OEM
Fast Thermal Conductivity Analyzer
Operating Manual
1.09KD180125MPL1V05© MessKonzept GmbH
Page 1: ...Filename FTC200_OEM_Englisch_1 09KD180125MPL1V05 docx Seite 1 von 10 FTC200 OEM Fast Thermal Conductivity Analyzer Operating Manual 1 09KD180125MPL1V05 MessKonzept GmbH ...
Page 2: ... 2 Description 4 3 EMC 5 4 Measuring gases and ranges 5 5 Scope of Delivery 5 6 Electrical Connection 6 7 Pneumatic Connection 7 8 Mounting 7 8 1 Thermal Insolation 8 9 Communication with the FTC200 OEM 8 9 1 Remote Control via Serial Communication 8 9 2 SetApp2 0 9 10 Specifications 10 ...
Page 3: ...e measures Explosive gases may not be lead in the FTC Upon installation the protection class has to be considered The ambient atmosphere may not be corrosive OEM device with protection class IP00 demand thermal and electric insulation as well as mechanical protection for operation The specifications of the device and its manual have to be observed strictly Please fill out questionnaire 2 01 1FB180...
Page 4: ...ure depends on the individual gas components and on the composition of the mixture Under certain conditions therefore the concentration of individual gas components can be determined by measuring the thermal conductivity In other cases the identification of certain properties of gases is of special interest e g for quality control or safety reasons Often these properties are related to the thermal...
Page 5: ...8 100 He N2 or air 0 100 0 0 8 97 100 He Ar 0 100 0 0 5 98 100 CO2 N2 or air 0 100 0 3 96 100 CO2 Ar 0 60 0 10 Ar N2 or air 0 100 0 3 96 100 Ar CO2 40 100 80 100 CH4 N2 or air 0 100 0 2 96 100 CH4 Ar 0 100 0 1 5 97 100 O2 N2 0 100 0 15 85 100 O2 Ar 0 100 0 2 97 100 N2 Ar 0 100 0 3 97 100 N2 CO2 0 100 0 4 96 100 NH3 H2 0 100 0 5 95 100 CO H2 0 100 0 2 99 100 SF6 N2 or air 0 100 0 2 96 100 Other gas...
Page 6: ...ply heater 24Volts Pin 04 Heater GND Pin 05 GND Pin 12 5VDC 5Volts from internal voltage regulator Connector X6 Pin 07 TxD TTL Serial TxD TTL level Pin 08 RxD TTL Serial RxD TTL level Pin 04 GND Connector X7 Pin 01 AIn1 Analog input 0 10Volts 50kOhm Pin 02 GND Pin 03 AIn2 Analog input 0 10Volts 50kOhm Pin 04 GND Pin 05 AOut1 Analog output 0 10Volts RL 1kOhm Pin 07 AOut2 Analog output 0 10Volts RL ...
Page 7: ... maybe significant lower This fact might give reason to condensation at dew points well below 50 C With proper heated lines and insulated connectors a dew point up to 50 C is permissible Important Notice Condensate will destroy the sensing element immediately condensate and dust tolerant version on request Optimal flow is in between 60l h and 80l h If the flow is kept stable during measurement as ...
Page 8: ... used must be temperature resistant up to 120 C It may not cover the PCBs We recommend an insulation material that is heat resistant up to 170 C The device is equipped with a temperature fuse switching off at 110 C on the surface of the stainless steel body This device is specified for an ambient temperature at direct place of operation between 10 C to 50 C Please consider the waste heat of other ...
Page 9: ...for the manual Remote Control via Serial Communication 9 2 SetApp2 0 Messkonzept offers a software for the operation of our devices It enables monitoring of measuring values and managing the settings of the FTC devices Especially for the OEM series the SetApp2 0 makes calibration setting thresholds and other things much easier The SetApp2 0 can be found on our website in the download section 1 ...
Page 10: ...0min 1h for small ranges Flow rate 10l h 100l h recommended 60l h to 80l h T90 time 1sec at flow rate higher 60l h Noise 1 of smallest range Drift at zero point 2 of smallest range per week Repeatability 1 of range Error due to change of ambient temperature 1 of smallest range per 10 C Error due to change of flow at 70l h 1 of smallest range per 10l h Gas pressure absolute 2000kPa 20bar Error due ...