INITIAL INCUBATOR CONFIGURATION
The Smart-Well incubator must be properly configured prior to use. The following options affect the
overall operation of the incubator, and can be configured from the
Tools
and
Advanced Tools Menu
s
(See the
User Interface Reference
for detailed procedures).
1.
Set the
Language
.
The default language is English.
2.
Change the
Administrator Password
.
The Administrator password is required to access the
Advanced Tools Menu
and all
options therein. This password should be maintained per your organization’s policies.
3.
Set the
Incubation Time
.
Any Smart-Read BI that has not turned positive in this duration will be considered negative
by the incubator. The default value is 10hours.
US Healthcare Intended Use:
Smart-Read
EZTest is a Self-contained Biological Indicator (SCBI) intended for use in determining
the efficacy of steam sterilization processes. The SCBI may be used in the following steam
sterilization cycles:
Cycle Type
Cycle Temperature
Cycle Exposure Time
Gravity
121°C
30 minutes
Gravity
132°C
10 minutes
IUSS Gravity
132°C
3 minutes*
Pre-Vac
132°C
4 minutes
Pre-Vac
135°C
3 minutes
*Unwrapped nonporous devices only.
Smart-Read
EZTest has a validated reduced incubation time of 10 hours.
4.
Set the
Date
and
Time
.
Please note that the incubator will not automatically adjust for any daylight savings practices
observed in your region.
The incubator’s date and time cannot be adjusted while an incubation cycle is active.
5.
Adjust
Sound
volume (optional).
The incubator’s volume can be adjusted and muted. This setting affects all sounds produced
by the incubator.