background image

Series SBLTX Submersible Level Transducer

Specifications - Installation and Operating Instructions

Bulletin L-37

The SBLTX Submersible Level Transducer

is manufactured for years of trouble free

service. The transducer consists of a piezoresistive sensing element, encased in a 316

SS housing. Bullet nose design protects diaphragm from damage. Comes equipped

with a 270-pound tensile strength, shielded, vented cable. Ventilation tube in the cable

automatically compensates for changes in atmospheric pressure above the tank.

Intrinsic Safety Approval Classification

The SBLTX is UL listed for use in Hazardous (Classified) Locations. The protection

method is by Intrinsic Safety, “ia”.  It was investigated by UL under UL Standard 913

8th Edition, CAN/CSA C22.2 No. 60079-0:15 and CAN/CSA C22.2 No. 60079-11:14.

Hazardous (Classified) Location Intrinsically Safe For:

Class I Div. 1 Groups A,B,C,D

Class II Div. 1 Groups E,F,G

Class III Div. 1 

Class I Zone 0 AEx ia IIC T4 Ga

Zone 20 AEx ia IIIC T135°C Da

Ex ia IIC T4 Ga

Ex ia IIIC T135°C Da

Ta = -20°C to 80°C (ETFE Cable) 

Ta = -20°C to 65°C (Polyurethane Cable)

Install in accordance with Control Drawing 001833-43. 

See Control Drawing 001833-43 for Entity Parameters.

ATEX: EU TyPE CErTIFICATE NO. DEmkO 18 ATEX 2080

ATEX STANDArDS: EN 60079-0:2012/A11:2013

EN 60079-11:2012 

ATEX CLASSIFICATION:       0518      II 1 G Ex ia IIC T4 Ga (-20°C ≤ Tamb ≤ 80°C

(ETFE Cable)) (-20°C ≤ Tamb ≤ 65°C (Polyurethane Cable))

0518      II 1 D Ex ia IIIC T135°C Da (-20°C ≤ Tamb ≤ 80°C (ETFE Cable))

(-20°C ≤ Tamb ≤ 65°C (Polyurethane Cable)) 

IECEx CErTIFICATE OF CONFOrmITy: IECEx UL 18.0086

IECEx STANDArDS: IEC 60079-0: 2011 6TH ED. 

IEC 60079-11:2011 6TH ED.

IECEx CLASSIFICATION: Ex ia IIC T4 Ga (-20°C ≤ Tamb ≤ 80°C (ETFE Cable)) 

(-20°C ≤ Tamb ≤ 65°C (Polyurethane Cable))

Ex ia IIIC T135°C Da (-20°C ≤ Tamb ≤ 80°C (ETFE Cable)) (-20°C ≤ Tamb ≤ 65°C

(Polyurethane Cable))

INSTALL IN ACCOrDANCE WITH CONTrOL DrAWING 001833-46

SEE CONTrOL DrAWING 001833-46 FOr ENTITy PArAmETErS.

SPECIFICATIONS

Service: 

Compatible liquids.

Wetted  Materials: 

316  SS,  316L  SS,  epoxy  adhesive;  Cable:  Polyether

polyurethane or ETFE; Bullet Nose: PVC.

Accuracy:

±0.25% of full scale.

Temperature  Limit: 

ETFE  Cable  Equipped -4  to  176°F  (-20  to  80°C);

Polyurethane Cable Equipped -4 to 149°F (-20 to 65°C).

Compensated Temperature Range:  

-4 to 176°F (-20 to 80°C).

Thermal Effect: 

Less than ±0.02% full scale/ °F.

Pressure Limit: 

2X full scale.

Power Requirement: 

10 to 28 VDC.

Output Signal:

4 to 20 mA DC, 2-wire.

Response Time: 

50 ms.

Max. Loop Resistance: 

900 ohms.

Electrical Connections: 

Wire pigtail.

Mounting Orientation: 

Suspended in tank below level being measured.

Weight:

2.2 lb (1.0 kg).

Agency Approvals: 

CE, See Intrinsic Safety Approval Classification.

MERCOID DIVISION

DWYER INSTRUMENTS, INC.

P.O. BOX 258 • MICHIGAN CITY, IN 46360 U.S.A.

Phone: 219/879-8000              www.dwyer-inst.com

Fax: 219/872-9057                  e-mail: [email protected]

BREATHER TUBE

SUBMERSIBLE
CABLE

1/2 NPT

6-15/16

[176.21]

Ø1

[Ø25.40]

WARNING:

Use  with  approved  safety  barriers  using

entity evaluation.