IMPORTANT SAFETY INFORMATION

Checkpoints before use 

1.  Adjust the armrests to a suitable height and lock the armrest by turning the adjustment knob clockwise.

2.  Fit the backrest tube in the intended bracket and lock the adjustment knobs for the height adjustment of the backrest. 

3.  Fit the brake lever in the intended bracket (left or right side). Does not apply to electric brakes. 

4.  All adjustment knobs must be tightened before sitting in the chair. 

5.  The brake on the chair must always be applied when getting in and out of the chair.  

The brake must be checked on a regular basis.

6.  Charge the chair according to the enclosed instructions before using it for the first time (applies to electrical chairs).

7.  The chair must not be used with the back and/or armrests removed.

The REAL 9000 PLUS chair has replaceable components, accessories and functions with setting options for individual 

customisation. The chair must only be used by the person for whom, and for the purpose for which, it is intended. The 

chair is intended for indoor use only and must not be used outdoors. The chair must not be exposed to extreme cold 

or heat, prolonged sunlight or other radiation. The chair must not be exposed to water, other fluids or chemicals. 
Warning! Metal surfaces may become very hot if they are exposed to direct sunshine.  

Read this manual carefully before using the chair for the first time.

•  REAL 9000 PLUS can affect and be affected by 

equipment in its immediate vicinity that generates 

electromagnetic fields; e.g alarm systems, power 

generators. Mobile phones and similar equipment 

can also interfere with the control system. Mercado 

Medic AB recommends that such equipment should 

not be stored any closer than 3.3 m(~10 ft) from the 

chair.

•  REAL 9000 PLUS has an electric lifting mechanism 

that can lift a person high enough to cause a risk of 

falling. If a person cannot get out of the chair from 

its highest position, then the seat height should 

never be set to higher than the position at which 

your feet touch the ground. The only exception to 

this is cases where you can summon help via alarm 

or telephone.

•   NOTE! The electrical functions on REAL 9000 PLUS 

chairs may be used for maximum 10% of the 

time, maximum 2 minutes every 20 minutes. The 

electronics in the chair measure the activity and 

shut down the electrical functions if this limit is 

exceeded.

•  Only electrical equipment from Mercado Medic AB 

may be used on the REAL 9000 PLUS.

•  Warning! Do not plug any accessory, other than 

those supplied by Mercado Medic AB, into the 

control system’s contacts.

•  The chair must be operated with care. Before 

moving off or operating the chair, consideration 

must be given to the surroundings. Any pets or 

children in the area could be easily run into. The 

chair must only be used to sit on and should not be 

regarded as a toy by either children or adults.

•  REAL 9000 PLUS is approved for transportation/

storage at temperatures of between -25°C and 

+60°C(-13°F and 140°F) and up to 90% non-

condensing relative humidity.

•  Max user weight,  REAL 9000 PLUS 150 kg (330 lbs).
•  The REAL 9000 PLUS should not be used on any 

incline greater than 5 degrees. If the chair is used or 

the brake is applied at steeper inclines, it may begin 

to slide which may cause the chair to fall over. 

•  If shipped, the chair must be transported in the 

intended cardboard box on a pallet. 

•  If the chair is transported by car, it must be strapped 

in place using straps over both the frame and the 

seat, and it should be set to the lowest possible seat 

height.

•   The chair is tested and approved for use between 

+5°C and +40°C (41°F and 104°F)  and 15 to 90%, 

non-condensing humidity. If the chair has been 

stored in an environment outside these limits, the 

chair must be left to acclimatise before use in order 

to ensure that it can be used safely.

•  Repairs and other technical measures may only be 

carried out by personnel authorised by EMP Living 

LLC.

•  Maximum height 2000m(~6500 ft) above sea level.
•  The REAL 9000 PLUS must not be moved when the 

seat is in the raised position.

•  Expected service-life ten (10) years.
•  REAL 9000 PLUS can be supplemented with a 

variety of accessories. Only accessories supplied 

by Mercado Medic should be used. All accessories 

must be fitted so that there is no risk of small parts 

coming loose. All the cables must be secured with 

cable ties to minimise the risk for strangulation.

•  Mercado Medic armrests, seats, backrests 

and other parts intended for skin contact are 

in compliance with the EU Regulation (EC) No 

1907/2006 concerning the Registration, Evaluation, 

Authorisation and Restriction of Chemicals (REACH).

FAMILIARISE YOURSELF WITH YOUR REAL 9000 PLUS

Summary of Contents for REAL 9000 PLUS ADULT

Page 1: ...lls of products and accessories We reserve the right to make changes to this manual and its contents ADULT REAL 9100 PLUS CHILD MINI REAL 9400 PLUS COXIT REAL 9800 PLUS TABLE OF CONTENTS Assembly for...

Page 2: ...20 07437 01...

Page 3: ...mechanism 4 5 5 Standard backrest mechanism 5 5 4 3 6 6 For manual brake remove the chair from the cardboard box Remove the brake lever that is held in place with plastic wrap along the brake shaft a...

Page 4: ...maximum 2 minutes every 20 minutes The electronics in the chair measure the activity and shut down the electrical functions if this limit is exceeded Only electrical equipment from Mercado Medic AB ma...

Page 5: ...chanism should be checked regularly for dust dirt and stability Raise the seat to its highest position Clean with a cloth Do not use any water or solvents Then lubricate the cylinder using a thin shee...

Page 6: ...in workmanship or material The warranty only applies to the original purchase and terminates if you transfer modify or sell the product Warranty period Welded metal parts 10 years Gas springs and ele...

Page 7: ...l 1 Seat tilt Crank adjustment turn the knob 2 Seat tilt Gas spring pull the lever backwards 3 Seat tilt Gas spring pull the lever upwards 4 Seat tilt Electric pull the lever upwards downwards 5 Seat...

Page 8: ...laced here move with the seat tilt 2 Armrests placed here do not move with the seat tilt 3 There are 3 positions When position 3 is selected the chair must be equipped with base extensions 1 Forwards...

Page 9: ...hours In order to preserve the battery s service life it is important that it never becomes fully discharged as this can damage the battery cells If the chair is equipped with a hand control with cor...

Page 10: ...once a day The battery may only be changed by personnel authorised by Mercado Medic AB See assembly instructions Battery Replacement Troubleshooting for electrical functions on your chair There are n...

Page 11: ...20 07437 01...

Page 12: ...ler in North America EMP Living LLC 14941 NE 147th CT Woodinville WA 98072 Phone 425 949 7195 Email info empliving com Web www empliving com Mercado Medic AB Phone 46 0 8 555 143 00 14 Tryffelslingan...

Reviews: