6.4.3.3 Stability Parameters
Stability parameters for the controlled pressure can be found in the Control Settings App and can be
configured using the Stable Window and Stable Delay buttons. When the controller enters a stable condi-
tion the pressure indication color on the Home App will change from white to green. The Stable Window
button allows the user to enter a value as a percentage of the highest range transducer. This value repre-
sents the pressure window within which any setpoint value would be considered stable by the user. The
Stable Delay button lets the user add a desired delay until the pressure value is considered stable while
being in the stable window.
Stability
Parameters
Figure 6.4.3.3 – Stability Parameters
6.4.3.4 Control Volume
The Control Volume button in the Control Settings App allows the user to control pressure volume in
“cc”. The CPC4000 is capable of automatically identifying the control pressure volume and adjusting the
control parameters based on it. By default this button is set on “Auto”. Figure 6.4.3.4 displays the control
volume setting for the instrument.
Figure 6.4.3.4 – Control Volume