Operating Instructions - CPC4000
Industrial Pressure Controller
CPC4000
4.3 Approvals and Certificates
EC declaration of conformity
EMC directive
7)
EN 61326-1 emission (group 1, class A) and interference immunity (industrial application)
Low voltage directive
EN 61010-1
Certificate
Calibration
8)
Standard: A2LA calibration certificate (standard on factory)
Option: DKD/DAkkS calibration certificate
7)
Warning! This is class A equipment for emissions and is intended for use in industrial environments. In other environments, e.g. residential or commercial installations, it can intefere with
other equipment under certain conditions. In such circumstances the operator is expected to take the appropriate measures.
8)
Calibration in a horizontal position/operating position.
4.4 Working Ranges of the Controller Modules
Bi-directional or gauge pressure [bar (psi)]
1)
-1 (-15)
0
3.4 (50)
10 (150)
100 (1,500)
210 (3,045)
LPSVR MODULE ±0.17 bar (±2.5 psi)
2)
MPSVR MODULE ±0.35 bar (±5 psi)
2)
HPSVR MODULE -1 ... 5 bar (-15 ... +75 psi)
2)
EPSVR MODULE -1 ... 10 bar (-15 ... +150 psi)
2)
Absolute pressure [bar (psi)]
1)
0
4.4 (65)
11 (165)
101 (1,515)
211 (3,060)
LPSVR MODULE 0 ... 1 bar (0 ... 15 psi)
2)
MPSVR MODULE 0 ... 1 bar (0 ... 15 psi)
2)
HPSVR MODULE 0 ... 6 bar (0 ... +90 psi)
2)
EPSVR MODULE 0 ... 11 bar (0 ... 165 psi)
2)
1) Mixing of absolute pressure and gauge pressure transducers in a module is not possible.
2) Smallest acceptable transducer range
For controlling absolute pressure a vacuum pump connected at the exhaust port is required.