8.6 Automatic Contamination Prevention System
The Automatic Contamination Prevention System (Automatic CPS) is connected between the device
under test and the Measure/ Control port of the instrument by the backplate located on the rear panel
(Figure 8.6-B). The primary function of the Automatic CPS is to prevent liquid and particulate contami-
nents from entering into the instrument. The Automatic CPS (Figure 8.6-A) is designed to interface with
the CPC3050 and is energized to purge the device under test of any contaminents by an automated
purge sequence within the CPC3050 user interface.
Figure 8.6-A -
Automatic Contamination Prevention System
Automatic CPS Backplate
Figure 8.6-B - Rear
Panel with Automatic CPS Backplate
i
Notice
The system should only be operated by trained personnel who are familiar with
this manual and the operation of the instrument.
Warning
!
Warning! The pressure connections must be installed according to the following
instructions, observing the relevant regulations. The installation is to be per-
formed by trained, authorized personnel, knowledgeable in the safety regulations
for working on pneumatic/hydraulic systems