Operating Instructions - CPC3050
97
High-Speed Pressure Controller
CPC3050
8.4 Fittings
A variety of fittings are available at the time of order with a max range of ≤ 3,060 psi (211 bar), or the
CPC3050 can be ordered without adaptor fittings. Available fittings are shown in Section 9.2 - Spare
Parts. The barometric reference comes with a standard barb fitting.
8.5 Remote Calibration
A remote calibration sled, power supply, cable and software are available to calibrate the optional baro-
metric reference. A PC with an RS-232 connection is required. The kits is supplied with an instruction
manual. For instructions to remove the barometric reference, see section 9.3 Transducer Removal.
8.5.1 Barometric Reference Calibration Sled (CPX-A-C4-5)
The CPC3050 Calibration Sled Kit is available to provide a way to calibrate the barometric reference re-
motely. Calibration of the Barometric Reference can be performed remotely using the Cal sled, a PC and
the software provided.
Figure 8.5.1 – CPC3050 Barometric Reference Cal Sled Kit
8.5.2 External Calibration Procedures
To calibrate the barometric reference remotely, use the same setup described in section 10 of this manu-
al. The only difference is that the interface with the transducer is through the software provided and a PC,
not directly through the instrument.