Operating Instructions - CPC3050
93
High-Speed Pressure Controller
CPC3050
:DEFine[N] <”CCC”, f>
Sets the user defined units for index N. If N is not
sent, will default to user units 1. Requires the de-
sired unit name in double quotes, multiplication
factor from Pascal (base units is always Pascal).
Index N is 1 or 2
7.8.2 Error Messages
All remote commands sent to the CPC3050 are shown in the Trouble Shooting App, under the “Remote”
button. If there is a syntax error, an error message will appear below the errant command. Local errors
and remote command errors will appear under the “Error” button. Maximum of 100 errors are stored and
can be retrieved.
The error messages can also be seen remotely when “SYST:ERR?” is sent any error in the error queue is
returned, otherwise the following is returned: 0,“No Error”.
Error String Returned
0, “No error”
1,”Overflow” (occurs when remote interface buffer of commands has become full. Command will not be
processed)
102, “Syntax error”
102,”Syntax error;Unbalanced string.”
105,"Parameter not supported, parameter 1"
105,”Parameter not supported, parameter 2”
108,”Parameter not allowed”
109,”Missing parameter”
113,”Undefined header”
113,”Command Unknown”
114,”Parameter Out Of Range”
114,”Header suffix out of range”
131,”Invalid suffix;Unit suffix not defined.”
200,”Execution error;Query or command violation”
200,”Execution error;Type mismatched, parameter 1”
200,”Execution error;Type mismatched, parameter 2”
203,”Access error; Incorrect password”
220,”Parameter error”
240,”Hardware error
500,”<s>” (where <s> is the undefined error error string generated by the pneumatic module)
201,”Query only”
207,”Enumeric value not in union”
601,”Module not available”
602,”Sensor not available”
999,”<s>” (where <s> is the undefined (unique) error string)