Operating Instructions - CPC3050
85
High-Speed Pressure Controller
CPC3050
7.8 GE PACE (SCPI) Emulation Command Set
The CPC3050 will accept and respond to the SCPI remote commands of GE PACE 5000/6000 remote
commands using the option SCPI_GE. Section 11, Technical Assistance explains the process of en-
abling the SCPI GE command set. Table 7.8.1 shows a complete list of supported commands.
7.8.1 SCPI GE Supported Commands and Queries
Table 7.8.1 SCPI GE Supported Commands and Queries
Command
Response/Function
CALCulate[x]
[x]=x=1=Channel A
[:LIMit]
:LOWer<n>
Set the minimum control limit *
:LOWer?
Get the minimum control limit *
:UPPer<n>
Set the maximum control limit *
:UPPer?
Get the maximum control limit *
CALibration[x]
[x]=x=1=Channel A
[:PRESsure]
:POINt?
Returns the number of calibration points used in
the active sensor’s linearity correction. Calibra-
tion functionality must be enabled by sending the
calibration password. See SYST:PASS:CEN
:ZERO
:AUTO
Start autozero process
:AUTO?
Returns 1 if autozero process is running, 0 other-
wise
VALVe ON | OFF | 1 | 0
ON or 1 puts in vent mode, OFF or 0 puts in mea-
sure mode
:VALVe?
Returns 1 if in vent mode, 0 otherwise
INPut[x]
[x]=x=1=Channel A
INSTrument
[x]=x=1=Channel A
:CATalog[x]?
Returns the sensor pressure maximum in native
units, native pressure units, native pressure type.
Ex// INST:CAT?
:INST:CAT “1000.000psig”
:CATalog