Operating Instructions - CPC3050
55
High-Speed Pressure Controller
CPC3050
6.4.10.1
Preset Steps
The user can have up to 12 different points along the range of the device under test (DUT). The Preset
Steps button allows the user to enter the number of points and then automatically divides the range in
equivalent percentages. For example if the user needs 5 distinct points along the DUT range then enter 5
in the numeric keypad and press Enter [
] button. CPC3050 will create setpoints at 20%, 40%, 60%,
80% and 100% of the DUT range. The user also has the option of changing these percentage values as
per his/ her desire by clicking on the individual Percentage button on the Step Settings app. The user can
enter any percentage between 0 and 100. The desired value can then be chosen by checking the box
[ ] beside the Percentage button.
Figure 6.4.10.1-A – Preset Steps Entry
Figure 6.4.10.1-B – Percentage Value Entry