7 Sterilization
31
2.
Confirm with [ENTER]. Now you can display the
program run or information about your steam sterilizer
(e.g. serial number, device software version and
selected values).
Manual program abort
You can abort a current program in all phases. If you abort the program before drying begins, the load re-
mains
unsterile
.
WARNING
Hot steam can be released from the device when opening the door after a program abort.
This could result in scalding.
n
Use a tray lifter to remove the tray.
n
Never touch the sterile material, the sterilization chamber or the door with unprotected
hands. The components are hot.
Program abort before the start of drying
WARNING
Danger of infection from early program abort
Aborting a program before the drying phase begins means that the load is unsterile. This
endangers the health of your patients and practice team.
n
Re-pack the load if necessary.
n
Repeat the sterilization of the load.
Should you still wish to do so, proceed as follows to abort the program before drying:
1.
Press CANCEL on the action bar.