background image

6 Sterilization

14

Information regarding routine operation

The sterilization temperature before every sterilization procedure must amount to 180 °C. To this end, the
setting of the sterilization time must take into account the Equilibration time. The pure aborticide time
amounts to 30 minutes at 180 °C.

1.

Wait 15 minutes after the thermometer has reached 180 °C.

2.

Set the required sterilization time. Comply with the specifications under 

Selecting the sterilization

times

 [

}

 page 14] .

3.

Check the ventilator noise: ensure that the mechanical air movement is functioning correctly.

Setting the temperature

Set the temperature for the hot air sterilizer as follows:

1.

Turn the temperature control to 180 °C.

2.

Wait until the thermometer reaches 180 °C and allow the device to heat up for 15 minutes.

Ê

During this time, the hot air sterilizer heats up the complete sterilization chamber.

3.

Set the sterilization time via the “On/Off switch”. Comply with the information of the sterilization times
table.

Selecting the sterilization times

Do not exceed the specified loading quantities. Use only trays and cartridges made of aluminium. When
needing to pack with foil, use aluminium foil.

Table 1: Load type and operating time

load

Packaging

Pre-heating time Operating time

max. of 500 g, including 2 trays

unwrapped

15 min

60 min

max. of 2 kg, including 2 trays

unwrapped

15 min

75 min

max. 500 g inc. 1 standard tray cassette

wrapped

15 min

75 min

max. 2 kg inc. 1 standard tray cassette

wrapped

15 min

120 min

Removing the sterilized equipment

CAUTION

Danger of burns from hot metal surfaces

n

Allow the device to cool sufficiently before opening.

n

Do not touch any hot metal parts.

CAUTION

Unsterile instruments resulting from damaged or burst packaging. This endangers the
health of your patients and practice team.

n

Should the packaging be damaged or have burst, re-pack the sterilization material and re-
sterilize it.

Comply with the following when removing the sterilized equipment:

u

Use a tray jack to remove the tray.

u

Never touch the sterilized equipment, the device interior or the inside of the the door with unprotected
hands. The components are hot.

Summary of Contents for Sterilizer 75

Page 1: ...shed MELAG as the world s leading manufacturer in the instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing compe...

Page 2: ......

Page 3: ...air sterilizer on and off again 11 Opening and closing the door 12 Changing the mount 12 6 Sterilization 13 Preparing the sterilization material 13 Loading the dry heat sterilizer 13 Information regar...

Page 4: ...to the device n The power cable may not be replaced by a cable determined to be insufficient Danger of short circuit n Liquids may not be permitted to reach the interior of the device This could resu...

Page 5: ...perating Manual near to the device It represents a component of the product Formatting rules Example Explanation see Chapter 2 Reference to another text section within this document Symbols used Symbo...

Page 6: ...the danger of burns In affixing the CE mark the manufacturer declares that this product fulfils the basic requirements of the EMC and low voltage directive The device may not be disposed as domestic...

Page 7: ...ack Standard tray jack Standard tray cassette unperforated Intended use This hot air sterilizer is intended for use in the cosmetics sector the cosmetic chiropody and the veterinary sector It was desi...

Page 8: ...amp heating 3 Thermometer 4 Indicator lamp network 5 On Off switch and time setting 6 Door handle 7 Door opens forwards 8 Device foot 9 11 10 Fig 2 Rear view of the device 9 Fan 10 Connection for the...

Page 9: ...loading versions with devices with trays and standard tray cassettes are possible in the following siz es Tray WxHxD 19 x 2 x 29 cm Standard tray cassette WxHxD 19 x 4 x 29 cm Mounting 1 Mount 1 belo...

Page 10: ...clearance of 10 cm to other devices and walls especially flammable parts u Ensure sufficient ventilation Ensure sufficient clearance to the top for free removal of the warm air The device may not be u...

Page 11: ...wise The control lights network and heating will remain illuminated until the time expires The hot air sterilizer can be set to continuous operation In this case the hot air sterilizer must be switche...

Page 12: ...cedure has started this could result in cooling and the object not being steri lized for a sufficient time 1 Move the On Off switch to the 0 position to switch off 2 Open the door forwards to open 3 M...

Page 13: ...tion material together in blocks as this will prevent heat equalization u Do not place the sterilization material on cellulose as this can result in a heat build up u Always pack the sterilization mat...

Page 14: ...zation times table Selecting the sterilization times Do not exceed the specified loading quantities Use only trays and cartridges made of aluminium When needing to pack with foil use aluminium foil Ta...

Page 15: ...a specified below for the storage of sterilized equip ment u Comply with the maximum storage duration in accordance with the packaging type u Do not store the sterilized equipment in the decontaminati...

Page 16: ...ng bio indicators thermocouples or maximum thermometers Comply with the valid regional legal specifications When performing the spore test using biological indicators the spore packets are to be clamp...

Page 17: ...o remove the impurities with this method Only if the chamber mount or chamber seal face has persistent soiling should you use a mild stainless steel cleaning agent with a pH value between 5 and 8 To c...

Page 18: ...temperature has been set too low The temperature controller rotary knob was unintentionally turned or mal adjusted during unpacking or incorrect handling e g during cleaning Set a higher temperature T...

Page 19: ...ase observe the maximum load 2 kg including trays Maximum mains voltage supply variations 10 Maximum voltage range 207 253 V Relative humidity up to 31 C max 80 up to 40 C max 50 Degree of protection...

Page 20: ...cessories 20 11 Accessories Table 2 Accessories Designation of article art no Mounting 1 59030 Mounting 2 59040 Tray 02075 Standard tray cassette 00287 Tray jack 28890 Jack for standard tray cassette...

Page 21: ...e sterilization material to the required temperature of 180 C Sterilization material Unsterile sterilizable material which is still to be sterilized Sterilized equipment Also referred to as a batch a...

Page 22: ...0829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations BA_Hei luftsteril...

Reviews: