4 Setup and installation
10
4 Setup and installation
Requirements of the installation location
NOTICE
The mains plug and the fuse must be easily accessible.
n
Set up the device in such a fashion that the power plug can be disconnected from the mains
quickly following danger and that the fuse is easy to access.
u
Install the device in a dry and dust-protected location in the interior of a building.
u
Install the device with a minimum clearance of 10 cm to other devices and walls, especially flammable
objects.
u
Ensure sufficient ventilation. Ensure sufficient clearance to the top for free removal of the warm air.
The device may not be used as an installation device and not be used in the immediate treatment
area.
u
The installation surface must be level and have a load-carrying capacity of at least the weight of the
device.
Tests after set-up
Perform a temperature check after setting-up the device using thermo sensors or bio indicators. Place the
test equipment in the area of the sterilization chamber in which the sterilization temperature is reached the
most slowly. This also depends on the arrangement and nature of the sterilization material.