Malfunctions
If the steam sterilizer does not seem to be working properly (e.g. poor drying, warnings, or error reports) then
follow these instructions in order to exclude possible operational errors. Following these instructions continue
to work with the steam sterilizer. If the malfunction occurs repeatedly then contact our dealer, and authorised
MELAG customer service or contact MELAG directly. You should describe the problem precisely and include
the works number of your appliance.
9.1 Warning messages __________________________________________________
For the following warning messages, please observe the comments made and restart the program in question.
If the warning occurs repeatedly please consult your stockist.
Warning message
Cause / Remedy
Door not closed properly
•
Turn the door grip down until contact is made (display
should then show "Door closed")!
This message appears when there is too less of feed water in
the water storage tank
•
Check the level of water in the tank, and refill to max.
with distilled or demineralized water of the appropriate
quality
Conductivity of the demineralized or distilled water is above the
first limit value, a start is possible by pressing the "Start" button
once more:
•
Empty water from the water storage tank, clean tank with
distilled / demineralized water and refill to the MAX- mark
with purified water to specifications
Ac k now le dge
w ith but t on "-"
No S t art
pos si bl e
AT TE NTI O N!
Door ope n
AT TE NTI O N!
No f e e d w ater
Ac k now le dge
w ith but t on "-"
Fe e d w ater
r efill
No S t art
pos si bl e
Fe e d w ater
c he ck s uppl y
Fe e d w ater qua lit y
ba d