MELAG MELAseal 100 Economic Operating Manual Download Page 7

 

Operating manual MELAseal

®

100

 

Page 7 

Rev.: 0 – 99/0211 

BA_1_GB_seal100eco.doc 

5. German 

DIN 

regulations 

5.1. Width of sealed seams and clear space  ________________________________________ 

German Standard DIN 58953 Part 7 requires that 
sealed seams in transparent packaging material 
should have a width of at least 8 mm. With a width of 
10 mm for its sealed seams, the MELAseal

® 

100 is 

considerably better than this minimum requirement.  

DIN 58953 Part 7 specifies that a clear space of at 
least 30 mm be left below the sealing border 
between the sterilisation goods and the seam in 
order to ensure a perfect seal without wrinkles. This 
applies both for bags and transparent material from 
the roll.  

5.2. Storage  life  ________________________________________________________________ 

Sterile goods packaging  

Type of packaging  

Length of storage  

unprotected 

1

 

Length of storage 

protected 

2

 

Paper bags in accordance  
with DIN 58 953 Part 3 

Sterile goods single 

packaging 

24 hours  

6 weeks  

Transparent sterilization packaging 
in accordance with  
DIN 58953 Part 4

 

 

Two-stage sterile 

goods packaging 

6 weeks  

6 months  

 Sterile-goods 

store 

packaging 

3 years 

3,4

 

1) 

E.g. on shelves  

2) 

E.g. in cupboards or drawers  

3) 

Before opening sterile-goods store packaging, dust shall be removed from it in accordance with 
regulations.  

4)  

The storage life of sterile-goods store packaging which has been opened is not altered if it has been 
closed again in accordance with regulations.  

 

A comment on the table in Section 8.2 of the DIN 
standard says: 

"The information in the table only represents 
standard values. These do not mean that longer 
storage periods are not possible under particularly 
favourable storage conditions, or that a longer 
duration of storage necessarily leads to the pack 
contents being recontaminated during storage, but 
signify that if these periods are observed, 
recontamination of the contents when they are 
opened and correctly removed is extremely 
unlikely."  

An additional explanatory note goes into further 
detail: 

"The risk in storing sterile goods is not that micro-
organisms may penetrate the sterile goods 
packaging, provided that dry storage conditions 
are ensured and the packaging material has the 
required germ-proof qualities, but that when the 
sterile goods packaging is opened after storage, 
micro-organisms contained in dust particles may 
be stirred up and could recontaminate the sterile 
objects when these are removed.  

"The extent of dust loading depends upon the 
dust content of the air and on the duration of 
storage. For these reasons the storage period 
should be kept as short as possible."

 

5.3. VDE  regulations ____________________________________________________________ 

In accordance with current VDE regulations, this appliance is not suitable for use in areas at risk of 
explosion. 

It should only be serviced by the manufacturer or a specially authorised agent (dealer or customer service). 

Summary of Contents for MELAseal 100 Economic

Page 1: ...zation equipment for medical practice During this period MELAG has succeeded in becoming a leading manufacturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testi...

Page 2: ...rking with packaging rolls 4 3 Accessories 5 3 1 Work platform 5 3 2 Unit mounted roll holder Comfort 5 3 3 Wall mounted roll holder 5 3 4 Acrylic hood 5 3 5 Standard roll holder 5 4 Servicing 6 4 1 E...

Page 3: ...adjusted by turning the temperature knob 8 With the transparent packaging material such as MELAfol most commonly used in autoclaves by doctors and dentists one side paper one side foil as described in...

Page 4: ...ning the lever to the vertical starting position and sealing can be repeated as required if the main switsch is not switch off 2 2 Packaging cutter ____________________________________________________...

Page 5: ...the MELAseal Package Sealing Device is placed against the wall or if it is mounted directly on the wall then it is best to have a wall mounted roll holder positioned directly above it The wall mounte...

Page 6: ...blade _______________________________________ Note Let the appliance cool down first to avoid risks of burns Then disconnect the appliance from the mains power supply This is important for all repair...

Page 7: ...aging which has been opened is not altered if it has been closed again in accordance with regulations A comment on the table in Section 8 2 of the DIN standard says The information in the table only r...

Reviews: