background image

 

 

Technical details are subject to modification without prior notice 

 
 
 
 

Operating Manual

 

 
 

Reverse Osmosis

 

Water Purifier

 

 
 

MELAdem

®

55

 

 

 

 

 

 

 

 

 

 

Dear Doctor, 

Thank you for placing your trust in this reverse osmosis water purifier. 

Since 1951,  MELAG has  specialised  in  the  production  of  sterilisation  equipment  for  medical 
practice. During this period, our family-owned and operated company has succeeded in becoming 
a leading manufacturer of sterilisation equipment. More than 450,000 MELAG appliances world-
wide testify to the exceptional quality of our sterilizers. 

Before using the 

MELAdem

 

55 

please study this operating manual carefully. 

In  order  to  ensure  a  long  working  life  for  your  reverse  osmosis  water  purifier  and  to  preserve  its 
value, please make sure that it is serviced regularly. 

 

MELAG management and personnel. 

 

 

Summary of Contents for MELAdem 55

Page 1: ...practice During this period our family owned and operated company has succeeded in becoming a leading manufacturer of sterilisation equipment More than 450 000 MELAG appliances world wide testify to...

Page 2: ...onnection of the MELAdem system to a Cliniklav 25 autoclave 6 4 2 Connection of the MELAdem system to an autoclave Type 25 7 5 Maintenance instructions 8 5 1 Replacement of the Pre filter 8 5 2 Replac...

Page 3: ...utoclave the permeate water from the reverse osmosis module is also passed through an ion exchange unit This contains an ion exchange resin which is able to reduce any residual mineral contents of the...

Page 4: ...ner element should be renewed when it is exhausted 5 Reverse osmosis module The reverse osmosis module is the key element in the water purifier It consists of a thin wound polyamide membrane In additi...

Page 5: ...ose end in an outflow floor drain or a bucket 9 Unscrew the filter housing of the subsequent ion exchanger 4 remove the mixed bed resin cartridge and screw on the empty filter housing to flush the uni...

Page 6: ...ed water for storage tank 17 Water purifier reverse osmosis system 18 Storage tank for water purification system The MELAdem 55 reverse osmosis system can be directly connected to autoclaves that have...

Page 7: ...ter shortage The MELAdem 55 reverse osmosis system can be directly connected to autoclaves that have been prepared for connection of a non pressurized automatic water feed system With such a configura...

Page 8: ...ry before or after periods of inoperation Period of inoperation Precautionary measures Up to 2 weeks Turn off water intake More than 2 weeks Turn off water intake empty storage container On resuming o...

Page 9: ...resh resin the contents of one package knocking the element to compact the resin so that the entire contents can be included Close the lid of the container element Take the sealing ring out of the cyl...

Page 10: ...Water pressure min max 2 bar to 6 bar Water temperature min max 5 C to 35 C Water pH value min max 4 0 to 10 0 Total mineral content of mains water max 1500 mg l Iron content of mains water max 0 1 mg...