Cleaning and maintenance
Inspect the mount
on a weekly basis
for soiling, deposits or damage.
If you find any impurities, remove the trays or cassettes from the chamber from the front. Clean the soiled components.
When cleaning the mount for loading, trays etc. comply with the following:
n
Ensure that the sterilization chamber is not hot.
n
Use a soft, non-fuzzing cloth.
n
First soak the cloth with the cleaning alcohol or spirit and attempt to wipe away impurities.
n
Use a chlorine- and vinegar-free cleaning fluid.
n
A mild, non-scouring, stainless steel cleaning agent, with a pH value between 5 and 8 should only be used for
persistent soiling.
n
Remove all cleaning fluid residue after cleaning the mount.
n
Do not use any hard objects such as a metal saucepan cleaner or a steel brush.
Replacing the slide clips
Proceed as follows to replace any slide clips which show signs of wear:
1.
Remove the damaged slide clip by levering it upwards carefully
using a small slotted screwdriver. Ensure that the wires do not suffer
kinking or crushing.
2.
Press the new slide clip into the same position on the wire cross on
the metal bar until it snaps into position.