Preface
1.1
Scope
These operating instructions apply to the Universal vessel hold with TopCut H urine and
secretion bag cutter.
The TopCut H is fully integrated into the 2013 model Topline and Topic cleaning and
disinfection machines if the machine is to cut open and empty urine and secretion bags
as well as clean urine bottles, bedpans and commode pans.
TopCut is not to be used in appliances with an automatic door. Due to the increased
operating forces caused by the cutting mechanism and the weight of the filled bags, this
could cause damage to/malfunction of the door drive or the TopCut itself. MEIKO will
not assume any warranty in such cases.
1.2
General information
Please read these operating instructions carefully.
We accept no responsibility for damage caused by incorrect operation/failure to observe
the operating instructions. Modifications to the TopCut, and in particular, technical
modifications carried out without the manufacturer’s written authorisation, or any
modifications carried out by unauthorised persons, will lead to the complete loss of any
guarantee claims and will invalidate any liability for the product.
MEIKO is constantly working on the further development of all its products. Please
understand that we reserve the right to make changes to the scope of supply
concerning the design, equipment and technical features at any time. No claims may
therefore be based on the details, the images or the descriptions contained in these
operating instructions.
Reproduction or translation of this document, or parts thereof, is not permitted without
written permission from MEIKO. All rights in keeping with German copyright law are
reserved by MEIKO.
These operating instructions are provided free of charge with the TopCut.
Please look after them carefully as we will have to charge you for each additional copy
you require to cover our own costs.
2
Intended use
TopCut H is a universal vessel holder for the Topline cleaning and disinfection machine.
It empties, washes and disinfects urine bottles, bedpans and commode pans, as well as
hygienically cutting, emptying and disinfecting urine and secretion bags (see section on
"Suitable bag types") thus reducing the level and volume of waste.
Only bags of urine and low viscosity secretions which are allowed to be discarded with
the waste water may be cut open and emptied using the TopCut. Bags of viscous
secretions, such as coagulated blood, do not drain freely from the bag once it has been
cut open and are thus not suitable for use with the TopCut.
Bags of urine and low viscosity secretions may be disposed of in this way if they have
been classified suitable as part of the operator's risk assessment and taking into
account the disinfection performance of the cleaning and disinfection machine.