VNA0440
–
VNA0440e
VNA0460
–
VNA0460e
user manual V3.1
- 11 -
S-Parameters
The measured DUT characteristics
impedance
and
gain
are called
Scatter Parameters
or
S-Parameters
. They are written as ‘
Sxy
’, where x and y are port numbers. If the port
numbers x and y are the same, an impedance is implied. If the port numbers are different
it represents the gain or loss of a signal
from port y to port x
.
These are the S-parameters that the VNA04x0 can measure:
•
S11
: Impedance at port 1
•
S22
: Impedance at port 2
•
S12
: Gain or loss from port 2 to port 1
•
S21
: Gain or loss from port 1 to port 2
The VNA04x0e can also measure these parameters:
•
S13
: Gain or loss from port 3 to port 1
•
S23
: Gain or loss from port 3 to port 2
The impedance S-parameters can be converted to a number of formats such as complex
impedance, return loss, forward loss and SWR. These all represent the impedance.
The Gain S-parameters are usually represented as magnitude and phase. The phase can
also be represented as a
Group Delay
which is a dynamic delay characteristic of a multi-
port device.
Block diagram
ADC
ADC
DET
DET
ATT
BRIDGE
BIAS-T
BRIDGE
BIAS-T
BIAS-T
AMP
P1
P2
GEN
CPU
USB
GEN
ATT
MegiQ VNA04x0 Block Diagram
FILES