7 ALARM FUNCTiON
7
A
larm function
7.1 Description
The TORKEL alarm function is provided by a buzzer
and a relay connected to the <ALARM> - connector.
An external alarm device can be connected to this
connector if desired.
Devices higher than Cat ii must not be attached to the
alarm connector.
Relay contact
8 A / 28 V DC
0.28 A / 250 V DC (resistive load only)
8 A / 240 V AC
Connector insulation
Voltage to ground must not exceed 250 V.
The following events can cause an alarm to
be issued
▪
Warning level is passed.
▪
Discharging is stopped because a stop level is reached.
▪
TORKEL can not regulate the current to the desired level.
▪
Thermal protection device trips or a fan does not rotate as
expected.
▪
The connection to the battery is broken.
▪
Other fault situations such as battery voltage too high or
too low or excessive current through TORKEL.
Note
TORKEL will safely interrupt the discharge and
issue an alarm if any of the internal tempera-
tures should exceed the safe operating limits.
Resetting the alarm
1]
You can reset the alarm by pressing the infor-
mation bar.
Summary of Contents for TORKEL 910
Page 2: ......
Page 13: ...CS033663DE ZP CS01E TORKEL 900 series 13 4 Menu system...
Page 27: ...CS033663DE ZP CS01E TORKEL 900 series 27 6 External current measurement...
Page 31: ...CS033663DE ZP CS01E TORKEL 900 series 31 7 Alarm function...
Page 33: ...CS033663DE ZP CS01E TORKEL 900 series 33 8 How to obtain the desired current...
Page 41: ...CS033663DE ZP CS01E TORKEL 900 series 41 10 Troubleshooting...
Page 43: ...CS033663DE ZP CS01E TORKEL 900 series 43 11 Calibration...
Page 45: ...CS033663DE ZP CS01E TORKEL 900 series 45 12 Specifications...
Page 49: ...CS033663DE ZP CS01E TORKEL 900 series 49 12 Specifications...
Page 51: ......