3.1 Panel
2
3
4
1
8
9
7
6
5
1.
R1 SENSE
R2 SENSE
Voltage input
2.
CURRENT OUTPUT
1 mA - 6 A, 48 VDC
Current output
3.
R2 SENSE
Switch for selection of single/dual channel meas-
urement. Dual channel measurement is active
when the switch is in the “ON”position.
4.
Touchscreen
4 inch, backlit, monochrome display
5.
DISCHARGE
Built-in discharge circuit safely discharges the test
object when test is completed. LED and audio sig-
nal indicate the charging status.
6.
100-240 V ~ 5 A
50 / 60 Hz
Mains input and fuse box: Always use the power
cord supplied with the unit. The unit is powered
when the switch for power input is in the “on”
position.
7.
RANGE
Selector for test currents.
8.
ON/OFF
Toggle switch ON / OFF for generation of test cur-
rents. Red Indicator lamp lit during generation.
9.
Protective conductor terminal: To be connected to
station earth (ground) using separate ground lead,
see "2.2 Safety instructions" on page 8.