9. Settings
In this section various user settings can be adjusted or accessed.
9.1 General settings
Switch the rotary switch to the settings mode
.
Go to the general settings tab
Megger DLR02 icons
µΩ
µΩ
mΩ
.
Use soft key 2 to select required setting as listed below.
General settings
Key Press Beep
Can be set on or off.
Results Beep
Can be set on or off.
Background Colour
Toggle between yellow and white using the tick button.
Backlight Timer
Use soft keys 2 and 3 to select from off up to 60 seconds (default is 20 seconds).
Battery Technology
Toggle between NiMH and alkaline using the tick button.
Difference Meter
Toggle ON and OFF using the tick button.
Sleep Timer
Press the tick button and navigate using soft keys 2 and 3. Use the tick button to
set.
Restore Factory Settings
Select using the TICK button (
) . When the prompt screen appears, press and
hold the TICK button (
) for 3 seconds.
Calibration
Not user accessible.
Settings
www.megger.com
DLRO2
28